• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS Back to Search Results
Model Number FD-412LR
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device referenced in this report has not yet been returned to olympus for evaluation.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event or Problem Description
During an endoscopic submucosal dissection, the subject device was used.During the procedure, the doctor used the subject device for hemostasis, but the subject device did not activate output.The intended procedure was completed with another device.There was no patient injury reported.
 
Additional Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The subject device worked.There were no defects on the subject device.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the event occurred due to any of the following possible causes.The contact condition between the patient and the patient plate was bad.The subject device was not connected to the cord or the cord was not connected to the power supply correctly.The target area was immersed in blood or water.The current density decreased due to increasing of contact area between the water and the subject device.The instruction manual of the device has described the method of connection of a code.The instruction manual of the device has already warned as follows.Pulling the tissue when applying the current.
 
Additional Manufacturer Narrative
This supplemental report is submitting to correct "device product code".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS
Common Device Name
SINGLE USE ELECTROSURGICAL HEMOSTATIC FORCEPS
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key7600282
Report Number8010047-2018-01147
Device Sequence Number1501301
Product Code KNS
UDI-Device Identifier04953170359804
UDI-Public04953170359804
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K062517
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberFD-412LR
Device Lot Number83K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/25/2018
Supplement Date Received by Manufacturer06/28/2018
02/20/2019
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date07/17/2018
03/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-