| Model Number |
FD-412LR |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device referenced in this report has not yet been returned to olympus for evaluation.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
|
| |
|
Event or Problem Description
|
|
During an endoscopic submucosal dissection, the subject device was used.During the procedure, the doctor used the subject device for hemostasis, but the subject device did not activate output.The intended procedure was completed with another device.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This is a supplemental report to provide additional information.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The subject device worked.There were no defects on the subject device.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the event occurred due to any of the following possible causes.The contact condition between the patient and the patient plate was bad.The subject device was not connected to the cord or the cord was not connected to the power supply correctly.The target area was immersed in blood or water.The current density decreased due to increasing of contact area between the water and the subject device.The instruction manual of the device has described the method of connection of a code.The instruction manual of the device has already warned as follows.Pulling the tissue when applying the current.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is submitting to correct "device product code".
|
| |
|