Model Number ZMA00
Medical Device Problem Codes Bent (1059); Scratched Material (3020)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.If explanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that during insertion into the eye, the intraocular lens zma00 +22.0 diopter had a bent haptic and the lens was scratched.The incision was enlarged to remove the lens.Another zma00 was implanted as a replacement during the same surgical procedure.No sutures were used and no patient injury was reported.No additional information was provided to johnson and johnson surgical vision, inc.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 6/26/2018.Device returned to manufacturer? yes.Device evaluation: the product was received in a lens case.The product is inspected by a qualified inspector using a microscope with a 12x magnification.It can be seen that the lens is cut.There is missing piece of the optic edge and the haptic is also damaged, most probably due to explant.The condition of the return sample does not suggest that damage was introduced during manufacturing.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7600514
Report Number9614546-2018-00583
Device Sequence Number13511059
Product Code MFK
UDI-Device Identifier05050474536845
UDI-Public(01)05050474536845(17)210405
Combination Product (Y/N)Y
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
P080010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/05/2021
Device Model NumberZMA00
Device Catalogue NumberZMA00U0220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/22/2018
Supplement Date Received by Manufacturer09/07/2018
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-