| Model Number |
R5201260 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to return to quest medical for evaluation.A follow-up medwatch will be submitted if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
A report was received regarding an alleged issue which occurred with the mps 2 console.The report states that the console displayed ec 225 and issued the following alarms: insufficient blood inlet and excessive blood inlet.No patient complications resulted from the alleged issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation showed fluid intrusion on the left pump optical encoder and back out of a pin.The source of the fluid intrusion is unknown.The encoder and pin were replaced and the console passed all operational verification tests.Quest medical will continue to monitor complaints for trend.
|
| |
|