US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 3973
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a sub total gastrectomy procedure, at the first firing, the reload was connected to the stapler, the surgeon tried to fire the device, however the plastic knob did not advance.A new reload was used and connected the same stapler, and the procedure was continued.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional information: evaluation summary: post market vigilance (pmv) led an evaluation of one device.The loading unit was received partially fired with no visual abnormalities.The sulu was loaded into a pmv representative instrument.The single use loading unit (sulu) and representative device were applied to the appropriate test media with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.Replication of the reported condition may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7601466
Report Number2647580-2018-02884
Device Sequence Number18900957
Product Code GDW
UDI-Device Identifier20884521063188
UDI-Public20884521063188
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885222
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number3973
Device Catalogue Number3973
Device Lot NumberP7E0779X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Initial Date Received by Manufacturer 05/29/2018
Supplement Date Received by Manufacturer07/02/2018
08/08/2018
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date07/31/2018
10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-