| Model Number |
3708660 |
| Medical Device Problem Codes |
Connection Problem (2900); Device Operates Differently Than Expected (2913); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a health care provider (hcp) via a manufacturer's representative (rep) regarding a patient who was im planted with an implantable neuro stimulator (ins).It was reported that the extension set screw/metal contact was bent, further stating that while the surgeon was attaching the lead to extension , the extension metal contact was moving.The surgeon thought the potential for lead damage or impedances issue could potentially occur.No environmental/patient factors that may have led to or contributed to the issue were reported.A new extension was during the procedure and the issue was resolved.No patient symptoms were reported.No further patient complications were reported/ anticipated as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the extension, sn (b)(4) found the distal end setscrew was missing.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|