| Model Number |
SL18P |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information regarding the device and incident has been requested.
|
| |
|
Event or Problem Description
|
|
Holes in the extension outside where the clamp is fixed.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the investigation performed and the inspection of the sample, it was concluded that the reported failure mode (holes in tubing) could have been caused by clamping the venous extension with the guidewire inside.The failure mode is considered not associated to the manufacturing processes.This family of devices has a 100% leak test as part of the manufacturing routine as a counter measure to detect this type of failure mode.
|
| |
|