MEDCOMP 8F X 18CM HEMO-CATH; HEMO-CATH LT Back to Search Results
Model Number SL18P
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Blood Loss (2597)
Date of Event 05/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information regarding the device and incident has been requested.
 
Event or Problem Description
Holes in the extension outside where the clamp is fixed.
 
Additional Manufacturer Narrative
Based on the investigation performed and the inspection of the sample, it was concluded that the reported failure mode (holes in tubing) could have been caused by clamping the venous extension with the guidewire inside.The failure mode is considered not associated to the manufacturing processes.This family of devices has a 100% leak test as part of the manufacturing routine as a counter measure to detect this type of failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8F X 18CM HEMO-CATH
Common Device Name
HEMO-CATH LT
Manufacturer (Section D)
MEDCOMP
1499 delp drive
harleysville PA 19438
MDR Report Key7601956
Report Number2518902-2018-00036
Device Sequence Number11237183
Product Code MSD
UDI-Device Identifier00884908045444
UDI-Public884908045444
Combination Product (Y/N)N
PMA/510(K) Number
K893439
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/22/2022
Device Model NumberSL18P
Device Catalogue NumberSL18P
Device Lot NumberMKAM100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer05/16/2018
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date01/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 MO
-
-