• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETER Back to Search Results
Catalog Number CRUOS106
Medical Device Problem Codes Material Separation (1562); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one crosser catheter was returned for evaluation.The returned catheter was noted to be bent along the distal portion of the device; however shipping could have contributed to this condition.The outer catheter and inner guidewire lumen were found to be separated from the distal tip of the device.Saline was flushed through the device; no leaks were observed along the length of the device except at the separated material.An in-house guidewire was inserted and retracted through the returned device without issue.Therefore, based on the returned sample condition, the investigation was confirmed for the separated material.The investigation was unconfirmed for the alleged leak and guidewire incompatibility issues.The root cause could not be determined based upon available information.It was unknown if patient and/or procedural factors contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during treatment of a tibial chronic total occlusion, the guidewire seemed tight and the recanalization catheter allegedly was difficult to advance.It was further reported that upon removal of the recanalization catheter, the catheter was allegedly sticking on to the wire and saline was leaking from the side wall of the catheter approximately 15 cm from the distal tip.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETER
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7602038
Report Number2020394-2018-00869
Device Sequence Number19307255
Product Code PDU
UDI-Device Identifier10801741125611
UDI-Public(01)10801741125611
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2020
Device Catalogue NumberCRUOS106
Device Lot NumberGFCQ3699
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2018
Initial Report FDA Received Date06/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Patient Weight54
-
-