| Model Number |
3275335F |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) of rebu1327 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the picc began to leak at the hub on the red extension after several months.The catheter was removed and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is confirmed; however the exact cause is unknown.One 5 fr dual lumen powerpicc ft was returned for evaluation.An initial visual observation showed use residue within the sample and some slight discoloration of the outer surface of the catheter.A split was observed in the extension leg tubing of the red lumen, just within the distal end of the luer connector.A microscopic observation revealed the fracture surface of the split was rough and exhibited cracks/fissures and beach marks, which are typical of material fatigue; however the exact cause of the split is unknown at this time.A lot history review (lhr) of rebu1327 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the picc began to leak at the hub on the red extension after several months.The catheter was removed and replaced.
|
| |
|
Search Alerts/Recalls
|