STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL 13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); Pupillary Block (2026); No Code Available (3191)
Date of Event 05/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Work order search- no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2 mm micl13.2, -15.5 diopter, implantable collamer lens into the patient's left eye (os) on (b)(4) 2018.On (b)(6) 2018 the lens was removed and exchanged with a shorter length lens due to observation of excessive vault, significant reduction of irido-corneal angles, pupil block and angle closure with elevated iop (intraocular pressure).Additional surgical intervention (yag pi) was performed on (b)(6) 2018.It was reported that at patients last follow up, vault and iop (intraocular pressure) were much improved.
 
Additional Manufacturer Narrative
Additional surgical intervention of enlargement of pi and additional of new pi (yag) performed on (b)(6) 2018.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7602293
Report Number2023826-2018-00912
Device Sequence Number233190
Product Code MTA
UDI-Device Identifier00841542103404
UDI-Public00841542103404
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/30/2019
Device Model NumberMICL 13.2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/16/2018
Supplement Date Received by Manufacturer08/13/2018
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 YR
-
-