|
On (b)(6) 2018, the patient had an initial endovascular aneurysm repair (evar) procedure.Three devices were implanted including a zenith flex aaa endovascular graft bifurcated main body and two zenith flex with spiral-z technology aaa endovascular graft iliac legs.At the end of the initial procedure the grafts were wide open, patent and had no kinks.On (b)(6) 2018, one week after the implant, the patient came to the hospital presenting with onset of bilateral lower extremity pain, numbness and motor loss.It was then determined that the entire graft system had thrombosed.The diagnosis was, aortic thrombosis of bilateral lower extremities and acute limb thrombosis.The physician advised that the patient was compliant with aspirin and anticoagulants.Secondary interventions were required including a right axial bi-femoral bypass, bilateral lower extremity four compartment fasciotomy, left femoral to above knee bypass, thrombectomy of the right superficial femoral artery, the posterior tibial artery and the anterior tibial artery.The patient is in critical condition requires medical management.Mfr.Report # 1820334-2018-01604 filed to report thrombosed main body, mfr.Report # 1820334-2018-01605 filed to report thrombosed right leg graft, and mfr.Report # 1820334-2018-01635 filed to report thrombosed left leg graft.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Evaluation / investigation the device was not available for evaluation.Without the complaint device, a physical investigation was not able to be completed.Imaging was requested but not provided.A document based investigation has been performed.A review of device history records, instructions for use and quality control data was conducted.Each zenith device is shipped with the instructions for use (ifu), listing the indications for use, contraindications, warnings and precautions, and the correct deployment procedure.Per the ifu, vessels that are significantly calcified, occlusive, tortuous or thrombus-lined may preclude placement of the endovascular graft and/or may increase the risk of embolization.Pre-existing regions of stenosis/narrowing have been shown to increase the risk of a thromboembolic event (e.G., graft limb occlusion).Patient with poor outflow or a hypercoagulable state may be at an increased risk of a thromboembolic event.Excessive overlap 10 mm above the main body bifurcation may increase the risk of limb thrombosis.There is no medical history available that could shed light on the patient having thrombosis so soon after the procedure.There is no medication list or diagnoses for review.There is no allegation the device kinked, migrated or fractured.Thrombus formation is a known inherent risk with these devices.A review of the device history record for lot 8461398 found no non-conformances associated with this device.Thrombosis can occur for a variety of reasons, such as genetic predisposition, iliac tortuosity, external compression, stenosis, improper overlap, narrow inner iliac lumen, vessel damage, and rough surfaces.Tortuosity, compression, and pulsation can create high shear forces that stimulate thrombus growth.Where a small thrombus has formed, the clot can enlarge because the blood flow slows around the clot, allowing clot-forming elements to be deposited.Pulsation and a small inner lumen can occur through graft oversizing or undersizing.Based on the information provided in this complaint, patient condition, genetic predisposition and disease progression most likely contributed to this event.It was reported that the grafts were patent with no kinks after the initial procedure and that the patient was compliant with aspirin and anticoagulants.There was no additional medical history provided for this complaint and it was clarified that the patient did not suffer any abdominal or pelvic trauma.After review of the design history files and device history record, there is no evidence to suggest that the device was manufactured out of specifications.Imaging would aid in determining a definitive root cause, however genetic predisposition, patient condition and disease progression most likely contributed to this event.Based on the information available, the investigation conclusion is the cause cannot be traced to the device - adverse event related to patient condition, because all of the grafts implanted in the patient completely thrombosed 1-week post-procedure.At the end of the implant procedure, the grafts were patent with no kinks and the patient was compliant with aspirin and anticoagulants.Based on the information provided, the patient was likely in a hypercoagulable state.The ifu provided with the zsle device states, "patients with poor outflow or a hypercoagulable state (e.G., cancer) may be at an increased risk of a thromboembolic event." measures are in progress to address this failure mode.The appropriate internal personnel have been notified and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|