• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problems Break (1069); Component Falling (1105)
Patient Problem Hyperglycemia (1905)
Event Date 08/16/2017
Event Type  malfunction  
Manufacturer Narrative
The test reservoir does not lock in place and unable to perform the displacement test due to missing retainer and missing reservoir tube o-ring.The insulin pump was received with scratched case, pillowing keypad overlay, broken reservoir tube lip and minor scratched lcd window.
 
Event Description
The customer reported via phone call that their insulin pump was damaged.The customer¿s blood glucose level was 279mg/dl.Troubleshooting was performed for damage and noticed the o-ring had fallen out and is missing.Advised the device will need to be replaced.Advised to discontinue use of the insulin pump and revert to a back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7602709
MDR Text Key111456244
Report Number3004209178-2018-81653
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169656840
UDI-Public(01)00643169656840
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG1F876
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/16/2017
Initial Date FDA Received06/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age49 YR
Patient Weight302
-
-