MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY ARES ANTIBIOTIC-IMPREGNATED CATHETER KIT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY ARES ANTIBIOTIC-IMPREGNATED CATHETER KIT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 95001
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported there was difficulty implanting the ventricular catheter.It was stated the catheter was not able to be inserted into the ventricle of the patient after ¿many attempts.¿ the catheter was replaced and there was no injury to the patient.
 
Additional Manufacturer Narrative
Approximately 23 cm of the ventricular catheter was returned.The catheter was patent and met requirements for leak testing.Approximately 120 cm of the peritoneal catheter was returned.The catheter was patent and met requirements for leak testing.Proteinaceous debris was observed within the interior and exterior of the catheter.No other anomalies were noted.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARES ANTIBIOTIC-IMPREGNATED CATHETER KIT
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7602797
Report Number2021898-2018-00293
Device Sequence Number1383223
Product Code JXG
UDI-Device Identifier00643169762664
UDI-Public00643169762664
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K110560
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date08/14/2019
Device Model Number95001
Device Catalogue Number95001
Device Lot Number0008991598
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer06/19/2018
Initial Report FDA Received Date06/14/2018
Supplement Report FDA Received Date07/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-