• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL HA SOLID BACK 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT PSL HA SOLID BACK 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 540-11-50E
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Injury (2348); Osteolysis (2377); Inadequate Osseointegration (2646)
Date of Event 05/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: not performed as the device was not returned.Dimensional inspection not performed as the device was not returned.Functional inspection: not performed as the device was not returned.Material analysis: not performed as the device was not returned.Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported relevant discrepancies.There have been no other similar events for the lot referenced.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that patients' right hip was revised due to pain, osteolysis, and aseptic loosening.Intra-operatively, surgeon reported that patient had an acetabular discontinuity due to a healed fracture (rep reported that no plates or screws had been used to address the fracture).Patient was revised to competitor shell and liner with a stryker universal taper biolox head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL HA SOLID BACK 50MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7602932
Report Number0002249697-2018-01816
Device Sequence Number11111807
Product Code LPH
UDI-Device Identifier04546540237828
UDI-Public04546540237828
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2008
Device Catalogue Number540-11-50E
Device Lot Number96030301
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2018
Initial Report FDA Received Date06/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
Patient Weight87
-
-