| Model Number |
72201812 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device keeps moving during case.No back-up device available.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and various scratches and small scrapes were noted on the outside of the unit.A functional evaluation revealed that the unit had excessive oil within the bimba tube housing as well as on the bimba piston.The amplifier piston seal had failed and allowed oil to seep into the bimba housing.The unit also failed the 50 pound weight test.The piston migration was at 2.3 inches, which is indicative of hydraulic fluid loss.The complaint was confirmed and the root cause has been determined to be excessive oil loss caused by a bad seal within the amplifier piston.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
| |
|