• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK® 20 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Physio-control evaluated the customer's device and was able to duplicate the reported issue.Physio replaced the system controller pcb assembly, which resolved the reported issue.After other unrelated repairs were completed, proper device operation was observed.The device was returned to the customer for use.Physio-control further evaluated the removed system controller pcb assembly and determined the cause of the reported issue was due to integrated circuit, designator u61.The u61 was electrically shorted and caused the device to lock up and not complete the boot-up process.
 
Event Description
The customer contacted physio-control to report that their device would not complete the boot-up process intermittently.When the device would complete the boot-up process, the screen would start flashing.As a result, defibrillation may not be possible, if it were necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK® 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
megan marseglia
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key7603284
MDR Text Key111152154
Report Number0003015876-2018-00938
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-