| Model Number |
480322-07 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The instrument was returned for evaluation.Failure analysis could not confirm the customer reported failure.The instrument was placed and driven on an in-house system.The instrument passed the recognition and engagement tests.The instrument moved intuitively with full range of motion in all directions.The grips opened and closed properly.The instrument successfully passed the cut test and the energy deliver test.The grip force test was performed and measurements passed within specifications.The instrument was fully functional.A follow-up mdr will be submitted if additional information is received.Isi has reviewed the site¿s system logs with a procedure date of (b)(6) 2018.No related system errors were found to have occurred during the surgical procedure.Based on the information provided at this time, this complaint is being reported due to the following conclusion: during the da vinci-assisted surgical procedure, it was alleged that the vessel sealer instrument had an insufficient sealing issue.Post-operatively, the patient experienced bleeding and underwent another surgical procedure.During the procedure, cauterization was performed in the area where the vessel sealer instrument had been used.However, the root cause of the operative complication is unknown.The instrument was returned to isi, evaluated, and the customer reported failure mode could not be replicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted thymectomy procedure, the surgical staff observed bleeding while using the vessel sealer instrument "at the branched part of the right arm head vein." during the event, the audible tones indicating that the sealing cycle was complete, were reportedly heard.When the surgeon opened the jaws of the vessel sealer instrument, bleeding was confirmed.According to the initial reporter, "hemostasis was completed with bipolar soft coagulation of erbe vi0 dv." on an unspecified date/time post-operatively, the color of an unspecified drain was allegedly found to be "too thick (dark)".As a result, the patient underwent a vats procedure and the area where the vessel sealer instrument was used was cauterized with an unspecified bipolar device.
|
| |
|
Additional Manufacturer Narrative
|
|
On 14-jun-2018, intuitive surgical, inc.(isi) received the following additional information regarding the reported event: the vessel involved with the operative complication was a ¿branch part of the brachiocephalic vein.¿ the vessel was less than 2mm in diameter.In relation to the sealing cycle while using the endowrist vessel sealer instrument, ¿smoke¿ or tissue effect was observed.There was no tissue tension.It is unknown if the vessel was calcified.When the alleged sealing issue occurred, the surgeon had opened the jaws of the instrument and was reportedly ¿surprised¿ to find bleeding.The surgeon claimed that it had taken longer than usual for sealing.The site provided a video clip of the reported event which was reviewed by an isi clinical development engineer (cde).The video clip shows an attempted seal of a target vessel using the vessel sealer instrument.During the attempted seal, no tissue effect is observed and minimal smoke is emitted from the area.No auditory feedback in the video clip can be heard indicating that the sealing cycle was completed.After the cut function of the vessel sealer instrument is activated with no error messages/faults, the instrument's jaws are opened and bleeding is immediately observed.It appeared as though the targeted tissue showed no signs of sealing or desiccation.The bleeding was controlled and eventually stopped with bipolar energy from a fenestrated bipolar forceps instrument.Based on the current information provided, this complaint will remain reportable due to the following conclusion: during the da vinci-assisted surgical procedure, it was alleged that the vessel sealer instrument had an insufficient sealing issue.Post-operatively, the patient experienced bleeding and underwent another surgical procedure.During the procedure, cauterization was performed in the area where the vessel sealer instrument had been used.However, the root cause of the operative complication is still unknown.The instrument was returned to isi, evaluated, and the customer reported failure mode could not be replicated.
|
| |
|
Search Alerts/Recalls
|
|
|