• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 06/18/2018
Event Type  Injury  
Event Description
The recipient's device will reportedly be explanted due to medical issues.Explant surgery is scheduled.The recipient will be reimplanted at a later date.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
delilah garcia
28515 westinghouse place
valencia, CA 91355
6613621741
MDR Report Key7604035
MDR Text Key111143875
Report Number3006556115-2018-00271
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Audiologist
Type of Report Initial,Followup,Followup,Followup
Report Date 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/31/2015
Device Model NumberCI-1500-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/14/2018
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received07/17/2018
08/15/2018
08/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
-
-