OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 12, LONG, STERILE, SINGLE USE, 12 PCS; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22037C
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical devices have not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical devices are returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of the bladder tumor (turbt) procedure, the loop wires at the distal end of three hf resection electrodes broke.It is unknown when exactly this damage occurred and whether a fragment fell inside the patient.No further information was provided but there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
Additional information: describe event or problem; device evaluation: the suspect medical device was returned to olympus for evaluation/investigation.The investigation confirmed that the loop wires at the distal end of three hf resection electrodes broke off.Furthermore, one of the loop wires showed discolorations (temper color), another loop wire showed material deposits.The cause of this damage and the breakage of the loop wires is most likely a contact with a conducting object and resulting short circuits.Therefore, this event/incident was attributed to use error.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot numbers of the hf resection electrodes without showing any abnormalities.The case will be closed from olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of the bladder tumor (turbt) procedure, the loop wires at the distal end of three hf resection electrodes broke and fell inside the patient.However, no fragment remained inside the patient since they were reportedly retrieved.No further information was provided but there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
Additional information: describe event or problem; device evaluation: the suspect medical device was returned to olympus for evaluation/investigation.The investigation confirmed that the loop wires at the distal end of three hf resection electrodes broke off.Furthermore, one of the loop wires showed discolorations (temper color), another loop wire showed material deposits.The cause of this damage and the breakage of the loop wires is most likely a contact with a conducting object and resulting short circuits.Therefore, this event/incident was attributed to use error.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot numbers of the hf resection electrodes without showing any abnormalities.The case will be closed from olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of the bladder tumor (turbt) procedure, the loop wires at the distal end of three hf resection electrodes broke and fell inside the patient.However, no fragment remained inside the patient since they were reportedly retrieved.No further information was provided but there was no report about an adverse event or patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 12, LONG, STERILE, SINGLE USE, 12 PCS
Common Device Name
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
MDR Report Key7604340
Report Number9610773-2018-00057
Device Sequence Number1623567
Product Code GCP
UDI-Device Identifier14042761051566
UDI-Public14042761051566
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K790071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Model NumberWA22037C
Device Catalogue NumberWA22037C
Device Lot NumberP16Y0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/01/2018
Supplement Date Received by Manufacturer08/15/2018
Initial Report FDA Received Date06/15/2018
Supplement Report FDA Received Date08/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-