• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 9560524
Device Problems Device Inoperable (1663); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was pre-operatively diagnosed with lumbar disc herniation and underwent micro endoscopic discectomy.During the surgery, the flexible arm could not be fixed firmly due to wear of thread.The product came in contact with the patient.No patient complications were reported as the result of the event.There was a delay of less than 60 minutes.
 
Manufacturer Narrative
Product analysis: visual and functional evaluation revealed that the tension wheel and the mating threads are all that were returned.The interfacing threads have been damaged from what appears to be torsional overload.The tightening of the flex arm can place high pressure on the threads that can cause them to be damaged.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Type of Device
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key7604454
MDR Text Key111147292
Report Number1030489-2018-00854
Device Sequence Number1
Product Code LXH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9560524
Device Lot NumberMY17D001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-