Model Number MMT-751NAL |
Device Problems
Air Leak (1008); No Display/Image (1183); Improper or Incorrect Procedure or Method (2017); Moisture or Humidity Problem (2986)
|
Patient Problem
Hyperglycemia (1905)
|
Event Date 05/19/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin pump screen was blank at morning and bubbles outside the reservoir.Customer's blood glucose level was high and it was 300 mg/dl.Customer reported that the insulin pump was exposed to moisture.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.Insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
The insulin pump passed displacement test, self test, off no power test and all operating currents tested within the specification range.The insulin pump was monitored and tested with no blank display noted.The insulin pump was received with corroded battery tube noted.No moisture damage on electronics and motor assembly noted.
|
|