GE HEALTHCARE FINLAND OY S/5 ADU CARESTATION; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Sticking (1597)
Patient Problem No Patient Involvement (2645)
Event Date 05/22/2018
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The adjustable pressure limit (apl) valve was replaced to resolve the reported issue.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that the adjustable pressure limit knob can't be adjusted off of zero.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S/5 ADU CARESTATION
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE HEALTHCARE FINLAND OY
kuortaneenkatu 2
helsinki
FI 
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7605061
MDR Text Key111163483
Report Number9610105-2018-00025
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/22/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/29/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-