Model Number 16-02-80
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 12/18/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Initially, it was believed that only one device was effected by this event.However, multiple heater-cooler cables were returned to the original equipment manufacturer for further investigation.During the evaluation, damage to the fuse point of the plug indicative of electrical charring was identified on an additional cord.Based on this information, it was determined that 2 serial numbers were affected by the reported issue.The device current is insufficient to cause the damage discovered.A measurement of the conductor resistance showed no deviation from the specification.It was determined that the cause of the damage was most likely a faulty socket with an increased contact resistance.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the cable of a heater-cooler system 3t showed significant traces of smoldering.The reported issue was identified during investigation by the original equipment manufacturer.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7605433
MDR Text Key111855357
Report Number9611109-2018-01047
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeAU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16-02-80
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received06/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-