MAUDE Adverse Event Report: DEPUY MITEK LLC US RIGIDFIX FEM2.7MM B/T/B XPIN; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US RIGIDFIX FEM2.7MM B/T/B XPIN; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Catalog Number 210127
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/28/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is being closed since after multiple attempts to retrieve the product, the device has not been returned for evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.If and when the device in question is received, this file will be reopened and its contents made to reflect the results of the evaluation.Udi: (b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
Customer reports the sleeve broke when surgeon was drilling.Surgeon reports that he followed the normal procedure, and he didn't apply any particular force on the device, so he suspects the metal was too weak.A new sleeve was taken from another box in order to carry out this procedure.No adverse reaction on patient came out.Additional information received from our affiliate on may 06, 2013 indicates that nothing fell into the patient, and the procedure was extended for ten (10) minutes due to the problem with the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIGIDFIX FEM2.7MM B/T/B XPIN
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7605482
Report Number1221934-2018-50860
Device Sequence Number10833738
Product Code MAI
UDI-Device Identifier10886705000883
UDI-Public10886705000883
Combination Product (Y/N)N
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/02/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2015
Device Catalogue Number210127
Device Lot Number3658814
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/02/2013
Initial Report FDA Received Date06/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-