Catalog Number RBY2C0415
Medical Device Problem Codes Detachment Of Device Component (1104); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the gastroduodenal artery (gda) using ruby coils.During the procedure, the physician successfully deployed and detached four ruby coils in the target location using a non-penumbra microcatheter.The physician then experienced resistance while advancing another ruby coil out of the tip of the microcatheter, after it was more than halfway into the vessel.Therefore, the ruby coil was retracted.While retracting, the physician experienced more resistance and, consequently, the ruby coil unintentionally detached and unraveled.The physician was unable to flush the detached ruby coil out of the microcatheter and into the target location and, therefore, the physician used a snare device to remove the detached ruby coil.The microcatheter was also removed, and the procedure was completed using a new microcatheter and a new ruby coil.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7606218
Report Number3005168196-2018-01195
Device Sequence Number19349563
Product Code HCG
UDI-Device Identifier00814548012858
UDI-Public00814548012858
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/18/2025
Device Catalogue NumberRBY2C0415
Device Lot NumberF80631
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/21/2018
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date06/15/2018
Supplement Report FDA Received Date08/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Date Report Sent to FDA01/01/2005
-
-