• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO RESTORE SENSOR; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 37714
Device Problems Failure to Deliver Energy (1211); Failure to Interrogate (1332); Improper or Incorrect Procedure or Method (2017); Low Battery (2584); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 05/17/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient with an implantable neurostimulator (ins) via a manufacturer representative.It was reported that the patient was attempting to recharge their ins when the stimulation stopped.It was reported that the ins was overdischarged.The patient was unable to charge the ins.The ins could not be interrogated with the clinician programmer, recharger, or patient programmer.A reset was performed and the patient was able to start using the ins.It was reported that the patient was scheduled for replacement surgery on (b)(6)2018.No further complications are anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORE SENSOR
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7606460
MDR Text Key111221083
Report Number3004209178-2018-13564
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00613994610430
UDI-Public00613994610430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2012
Device Model Number37714
Device Catalogue Number37714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/07/2018
Date Device Manufactured12/14/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient Weight61
-
-