| Model Number |
CRPLUS |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio further evaluated the device's analog pcb assembly and determined that the cause of the reported issue was a failed relay, designator k1, on the analog pcb assembly.The k1 relay did not toggle to transfer energy.The customer received a replacement device.
|
| |
|
Event or Problem Description
|
|
Physio-control was performing preventative maintenance on a customer's device and observed that the device would not deliver defibrillation energy.As a result, defibrillation therapy would not be possible, if it was needed.There was no patient use associated with the reported event.
|
| |
|