It was reported an angioplasty of the posterior tibial artery with a ipsilateral femoral access using a 6fr boston scientific sheath was performed.An approach® hydro st microwire guide with a rubicon support catheter was used for the procedure.(it was reported, the nominal pressure of the balloon is 8 atm and the rated ruptured pressure is 16).The balloon was insufflated at 14 atm and ruptured.The balloon was removed immediately.The balloon was purged with saline, was insufflated with contrast medium and saline using a boston scientific insufflator.Additional information was received on 22may2018: the reporter clarified the inflation pressure used was 14lp and the contrast to saline used was ultraist 300 with 50% mix of contrast and 50% saline.The lesion was located in the distal posterior tibial artery and there was no angulation.The balloon burst longitudinally.There was little calcification in the site of the lesion.The physician previously used another balloon from a competitor, it did not burst.
|
|
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed no visible damage to the balloon portion of the device.However, a leak was noted at the proximal end of the catheter, immediately distal to strain relief.Kinks were noted at distances of 11.8 cm, 28.1 cm, 40.2 cm, 69.3 cm, 87.4 cm and 105.3 cm from the proximal end of the device.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|