MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00028-120
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A manufacturer representative went on site to evaluate the device.Functional and visual testing were conducted.Testing revealed that the device's fuses had blown.The fuses were replaced and the system checkout was successful.No further action required.
 
Event or Problem Description
Medtronic received information regarding an imaging device.It was reported that the device won¿t turn on.The site further reported that the outlet had blown a fuse which caused the mobile view station¿s (mvs) battery to deplete.The device was plugged into a known good outlet and the issue was resolved.This was observed outside of a surgical procedure.No patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Common Device Name
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
MDR Report Key7607415
Report Number1723170-2018-02806
Device Sequence Number13801919
Product Code OXO
UDI-Device Identifier00643169845091
UDI-Public00643169845091
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K092564
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,user facility
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 06/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Radiologic Technologist
Device Model NumberBI-700-00028-120
Device Catalogue NumberBI70000028120
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2017
Initial Report FDA Received Date06/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-