| Model Number |
406124 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While removing the catheters post procedure from the groin, the fast-cath hemostasis introducer ripped.Part of the device was still in the patient but able to be removed because the catheter was still in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product evaluation: the reported event of "the fast-cath hemostasis introducer ripped" could not be confirmed.The results of the investigation are inconclusive since the device was not received in the product performance engineering lab for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.The fast-cath introducer instructions for use (ifu) cautions the user to not attempt to advance or withdraw guidewire if resistance is felt.Use fluoroscopy to determine cause.The fast-cath introducers instructions for use (ifu) recommends the user advance dilator/sheath assembly with a twisting motion to avoid damage to the sheath or vessel.
|
| |
|