| Model Number |
72203360S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the camera head image fades in and out when the cable is moved, it is possibly a wiring issue.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation was performed by smith & nephew and could confirm the customer complaint for the image fades in and out.A visual inspection was performed and showed the connecting end of the camera head to the control unit shows water spots indicating a leak path.This is caused by the cap not be tightened down properly during the cleaning process.No manufacturing related defects were observed.
|
| |
|