• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RFB CONTROL UNIT DYONICS 25; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RFB CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Model Number 7211010S3E
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/08/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was overpressurizing; therefore, it had an inconsistent pressure.The case was completed with the same device.No patient injuries were reported.
 
Additional Manufacturer Narrative
Device evaluation: the device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of over pressurization could not be reproduced.Product passed functional testing per factory test with no faults or errors.Product passed functional testing during 6 hour burn-in on wet station utilizing low and high pressure and flow settings.Raw and zero transducer readings were normal and well within specs during all functional tests.At no time during functional testing did pump over pressurize, fluctuate, or become intermittent.All functions perform as expected.Pressure and flow readings were normal throughout burn-in on wet station.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFB CONTROL UNIT DYONICS 25
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7608481
Report Number1643264-2018-00468
Device Sequence Number1623845
Product Code HRX
UDI-Device Identifier00885554030822
UDI-Public(01)00885554030822
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K051326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7211010S3E
Device Catalogue Number7211010F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2018
Initial Date Received by Manufacturer 06/11/2018
Supplement Date Received by Manufacturer07/24/2018
Initial Report FDA Received Date06/17/2018
Supplement Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-