• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION TECNIS SYMFONY; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION TECNIS SYMFONY; MULTIFOCAL IOLS Back to Search Results
Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017 and (b)(6) 2018.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that tecnis symfony multifocal intraocular lens (model zxr00, +24.0 diopter) was explanted from patient's left eye in secondary procedure.There was no incision enlargement, no vitrectomy.An intraocular lens from another manufacturer was used as replacement lens for the patient.Patient was doing well at discharge.No further information provided.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation? yes.Returned to manufacturer on: 8/1/2018.Device returned to manufacturer? yes.Device evaluation: the device was received at the manufacturing site for investigation.Visual inspection shows there was cut on the optic of the lens and the lens was dirty.The cut on the lens, most probably to make an explant possible.The lens is dirty and contaminated.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.Investigation of the return sample and the condition of the return sample do not suggest the dirty and cut was introduced during manufacturing.The complaint cannot be confirmed.Manufacturing record review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaint was received for this production order.Historical data analysis: the search in complaint history revealed no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION
santa ana CA
MDR Report Key7608746
Report Number9614546-2018-00588
Device Sequence Number2144036
Product Code POE
UDI-Device Identifier05050474579170
UDI-Public(01)05050474579170(17)211211
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date12/11/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 05/21/2018
Supplement Date Received by Manufacturer08/24/2018
Initial Report FDA Received Date06/18/2018
Supplement Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-