MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN SECTOR HA ACET CUP 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN SECTOR HA ACET CUP 52MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-12-052
Medical Device Problem Codes Use of Device Problem (1670); Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Fatigue (1849); Foreign Body Reaction (1868); Pain (1994); Tissue Damage (2104); Distress (2329); Discomfort (2330); Injury (2348); Osteolysis (2377); Inadequate Osseointegration (2646); Limited Mobility Of The Implanted Joint (2671); Not Applicable (3189)
Date of Event 08/25/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pinnacle litigation records received.Litigation alleges progressively worse pain, lack of mobility, emotional distress, metallosis and elevated metal levels of cobalt and chromium.Revision note reported, on the femoral side there was a significant degree of bone loss.There was a large black stained on mass on the lateral aspect of the greater trochanter, this was filled with metal debris.There was a pathologic fracture at the base of the greater trochanter.The acetabular component was grossly loose and there was pronounced metal staining of the remaining native acetabulum and surrounding soft tissues.There was almost complete loss of the anterior column and severe thinning of the posterior column.Radiograph shows some osteolysis with an area of lucency at the acetabular medial wall.Laboratory results for blood metal ion level shows above 7 ppb.Doi: (b)(6) 2008 : dor: (b)(6) 2017 (left hip).
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pfs alleges discomfort and fatigue.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. udi: (b)(4).
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.   additional narrative: added (patient code).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d10 and h6 (device code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR HA ACET CUP 52MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6103142063
MDR Report Key7609251
Report Number1818910-2018-62393
Device Sequence Number1513009
Product Code LPH
UDI-Device Identifier10603295009160
UDI-Public10603295009160
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K031495
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/13/2017
Device Model Number1217-12-052
Device Catalogue Number121712052
Device Lot NumberBJ4H51000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/24/2018
Supplement Date Received by Manufacturer06/29/2018
02/26/2019
09/02/2019
02/07/2022
02/25/2022
Initial Report FDA Received Date06/18/2018
Supplement Report FDA Received Date07/03/2018
03/07/2019
09/18/2019
02/09/2022
02/26/2022
Date Device Manufactured03/16/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +1.5; PINNACLE MTL INS NEUT36IDX52OD; SUMMIT DUOFIX TAP SZ4 STD OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
Patient Weight61 KG
-
-