• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HALYARD HEALTH MICROCUFF ENDOTRACHEAL TUBE FOR PEDIATRICS, 3.0 MM, ORAL/NASAL MAGILL; VAP MICROCUFF ENDOTRACHEAL TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HALYARD HEALTH MICROCUFF ENDOTRACHEAL TUBE FOR PEDIATRICS, 3.0 MM, ORAL/NASAL MAGILL; VAP MICROCUFF ENDOTRACHEAL TUBES Back to Search Results
Model Number 35111
Medical Device Problem Code Air Leak (1008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is not possible, as a valid lot number was not provided.All information reasonably known as of 14 jun 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.(b)(4).
 
Event or Problem Description
Halyard received a single report that referenced two different incidences, which were associated with separate units.This is the first of two reports.Refer to 3011270181-2018-00017 for the second report.It was reported that the cuff pressure dropped over time.The cuffs require re-inflation several times each shift.The hospital is using new ventilator machines that alarm when an air leak is detected.Per additional information received 8 jun 2018, the nurse reports the risk of injury to patients is low, as the ventilator alarm ensures that the nurses know when to re-inflate the cuff.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of 28 jun 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Additional Manufacturer Narrative
The used sample was returned for evaluation.The device was visually inspected and no damage to the device was noted.The inflation lines and insertion areas appeared to have adhesive residues.The balloon cuffs were inflated with air and no leakages were observed.The inflation line, balloon cuff, and pilot balloons were submerged in water, and no leakage was observed.The complaint could not be confirmed as reported.No root cause could be determined.All information reasonably known as of 14 sep 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a halyard health product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROCUFF ENDOTRACHEAL TUBE FOR PEDIATRICS, 3.0 MM, ORAL/NASAL MAGILL
Common Device Name
VAP MICROCUFF ENDOTRACHEAL TUBES
Manufacturer (Section D)
HALYARD HEALTH
5405 windward parkway
alpharetta GA 30004
MDR Report Key7609294
Report Number3011270181-2018-00016
Device Sequence Number2281972
Product Code BTR
UDI-Device Identifier10680651351118
UDI-Public10680651351118
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K080821
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Model Number35111
Device Catalogue Number991035111
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/07/2018
Supplement Date Received by Manufacturer06/18/2018
08/28/2018
Initial Report FDA Received Date06/18/2018
Supplement Report FDA Received Date06/29/2018
09/14/2018
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN VENTILATOR MACHINE
Patient Weight3
-
-