| Catalog Number |
B12XT |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Only event year is known: 2018.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
During robotic surgery, trocar used for scope is broken.The reason is that when the temperature rises due to the light and the material of the melted trocars is put in and out of the apparatus, it is presumed that a physical bump has occurred and broken, there were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # p91z1j.Device analysis: the analysis results found that the b12xt instrument was received with the sleeve melted.In addition, the tyvek was returned along with the instrument.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.The lot history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the packaging process.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|