STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 58MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 502-03-58F
Medical Device Problem Codes Corroded (1131); Loss of Osseointegration (2408); Material Deformation (2976)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Unspecified Infection (1930); Injury (2348); Inadequate Osseointegration (2646)
Date of Event 05/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported relevant discrepancies.There have been no other similar events for the lot referenced.There have been no other similar events for the sterile lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Hip revision - mdm liner and tritanium primary cup - metal/metal wear - removed components and inserted antibiotic spacer.Patient was revised due to deep infection.Patient also had a pelvic fracture and the shell was found to have migrated medially, through the medial wall.Corrosion on the back of the metal liner was also reported.
 
Additional Manufacturer Narrative
An event regarding infection, periprosthetic acetabular fracture, loosening, wear and corrosion involving a tritanium shell was reported.The event was not confirmed.Method & results: product evaluation and results: the device was not returned however, one photo of the explanted shell was provided.The shell shows some biological material on the outer surface and rim of the shell.There are some marks on the inner surface of the shell but it is unclear what these are.Medical records received and evaluation: not performed as no medical records were returned for review.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there were no other events for the lot and sterile lot provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as device return, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time as no devices and insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Hip revision - mdm liner and tritanium primary cup - metal/metal wear - removed components and inserted antibiotic spacer.Patient was revised due to deep infection.Patient also had a pelvic fracture and the shell was found to have migrated medially, through the medial wall.Corrosion on the back of the metal liner was also reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 58MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7611181
Report Number0002249697-2018-01857
Device Sequence Number13483697
Product Code LPH
UDI-Device Identifier07613327040159
UDI-Public07613327040159
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Catalogue Number502-03-58F
Device Lot NumberMNA72L
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/24/2018
Supplement Date Received by Manufacturer07/06/2018
Initial Report FDA Received Date06/18/2018
Supplement Report FDA Received Date08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-