• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-611
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 05/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding revision due to pain involving a triathlon insert was reported.The event was not confirmed.Method and results: device evaluation and results: the reported device was not returned however photographs were provided for review.The photographs shows a recently explanted insert with nothing remarkable to report.Medical records received and evaluation: a review of the provided medical records x-rays by a clinical consultant indicated: soft tissue imbalance in the extensor tendon mechanism of the knee contributed to lateral tilt/shift of the patellar device with overload on the remaining contact sections with poly wear and damage becoming evident causing anterior knee pain requiring revision surgery with exchange of the patellar device and tibial bearing exchange for access reasons to the knee.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the investigation concluded that: soft tissue imbalance in the extensor tendon mechanism of the knee contributed to lateral tilt/shift of the patellar device with overload on the remaining contact sections with poly wear and damage becoming evident causing anterior knee pain requiring revision surgery with exchange of the patellar device and tibial bearing exchange for access reasons to the knee.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
As reported: "pain in anterior knee.Poly swap while in knee.All available information submitted".Explant pictures show that the patellar implant suffered visible damage.Patient was revised to a 6/11 cr insert and an a35 x3 patella.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7611383
Report Number0002249697-2018-01862
Device Sequence Number18877939
Product Code MBH
UDI-Device Identifier07613327050592
UDI-Public07613327050592
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2014
Device Catalogue Number5530-G-611
Device Lot NumberMHHM8Y
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2018
Initial Report FDA Received Date06/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient Weight114
-
-