MAUDE Adverse Event Report: ZYNO MEDICAL LLC ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM; LARGE VOLUME INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZYNO MEDICAL LLC ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM; LARGE VOLUME INFUSION PUMP Back to Search Results
Model Number Z-800
Medical Device Problem Codes Inaccurate Flow Rate (1249); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zyno medical is waiting for the arrival of the device.
 
Event or Problem Description
A distributor reported an over-infusion issue to zyno medical on (b)(6) 2018.The end user reported to the distributor that the affected pump was set for an infusion for two hours but completed the infusion in only an hour.When the distributor received the pump, the pump was tested for 3 times, sometimes within specification and other times had a flow rate twice as fast.No patient injury was reported for this case.
 
Event or Problem Description
This is a follow-up for the initially filed mdr (3006575795-2018-00053).
 
Additional Manufacturer Narrative
The pump was tested on 06/22/2018 and met all specifications.The pump was found to have a flow rate accuracy of -1.93% which is within +/-5% specification.The reported flow rate issue was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM
Common Device Name
LARGE VOLUME INFUSION PUMP
Manufacturer (Section D)
ZYNO MEDICAL LLC
177 pine street
natick MA 01760
MDR Report Key7611768
Report Number3006575795-2018-00053
Device Sequence Number1638275
Product Code FRN
UDI-Device Identifier00814371020006
UDI-Public00814371020006
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K130690
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZ-800
Device Catalogue NumberZ-800
Device Lot Number20120105-SH
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2018
Supplement Date Received by Manufacturer06/25/2018
Initial Report FDA Received Date06/18/2018
Supplement Report FDA Received Date07/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-