The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
|
During preparation for a coil embolization procedure, the nurse noticed that a ruby coil embolization coil was out of its introducer sheath upon removal from the dispenser hoop.The physician then tried to push and pull the ruby coil but it was not successful.Therefore, the ruby coil was set aside and the procedure was completed using another coil.
|
Results: the pusher assembly was kinked along its length.The proximal pet lock was damaged.The pusher assembly was fractured approximately 5.0 cm from the proximal end and embolization coil was detached inside the introducer sheath.The introducer sheath was ovalized approximately 18.5 cm-19.0 cm, 27.0 cm, 40.0 cm, 46.5 cm, 53.0 cm, 75.5 cm and 78.0 cm from the proximal end of the sheath.The embolization coil had offset winds along in multiple locations within the introducer sheath.Conclusions: evaluation of the returned ruby coil revealed the pusher assembly was fractured and the embolization coil was detached.If the ruby coil is mishandled during removal from the packaging, the pusher assembly may become fractured allowing the embolization coil to detach.This is further supported by the embolization coil being detached within the introducer sheath.Further evaluation revealed the friction lock was damaged, the pusher assembly and introducer sheath were kinked, and the embolization coil had offset coil winds.This damage was likely incidental to the reported complaint.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|