Model Number MMT-551NAS |
Device Problems
Break (1069); Component Falling (1105)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/22/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that their insulin pump was damaged.The customer¿s blood glucose level was 84 mg/dl.The customer reported that the little reservoir ring was not look in place and has a little peace the lip ring had fallen off.Troubleshooting was performed for damage and noticed that the reservoir was falling off.Advised the pump will need to be replaced.Advised to discontinue use of the pump and revert to a backup plan.The insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
The insulin pump passed the displacement test.Pump was received with partial broken reservoir tube lip and missing reservoir tube o-ring noted.The p-cap locks into the reservoir compartment properly noted.
|
|