Model Number MMT-7020B
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 03/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they had a low blood glucose value of 48, 63 and 40 mg/dl.Customer declined to troubleshoot for their blood glucose levels.Sensor updating and change sensor were also reported.The customer was treated with carbohydrate.The insulin pump is not expected to be returned for analysis.
 
Additional Manufacturer Narrative
Inspected one opened/used sensor and performed continuity resistance test.Sensor passed per specification.Also, performed bicarbonate buffer test.Sensor passed per specifications with accurate readings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020B ENLITE3 US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key7612705
Report Number2032227-2018-06419
Device Sequence Number11311402
Product Code OZP
UDI-Device Identifier00643169704060
UDI-Public(01)00643169704060(017)20181028
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/28/2018
Device Model NumberMMT-7020B
Device Catalogue NumberMMT-7020B
Device Lot NumberJ287
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer08/07/2018
Initial Report FDA Received Date06/19/2018
Supplement Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age63 YR
Patient Weight167
-
-