• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US VERSALOK DEPLOYMENT GUN; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US VERSALOK DEPLOYMENT GUN; ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR Back to Search Results
Catalog Number 214710
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/21/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation summary: the complaint device is not being returned, therefore is unavailable for a physical evaluation.Furthermore, no lot numbers were supplied which precludes conducting a batch history review or a lot specific search in the complaints handling system.Therefore, we cannot determine what caused the user to experience the reported event, we cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).Depuy synthes has been informed that the lot number is not available.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
Our affiliate reported when that when trigger was pulled, the gun did not function properly due to a loosened screw.Another product was used to complete the case.There were no adverse consequences to the patient.Affiliate stated procedure time was extended thirty (30) minutes.Please see also medwatch report 1221934-2018-50757.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSALOK DEPLOYMENT GUN
Common Device Name
ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7612870
Report Number1221934-2018-51931
Device Sequence Number11333190
Product Code LXH
UDI-Device Identifier10886705004416
UDI-Public10886705004416
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/21/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214710
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/21/2013
Initial Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-