• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. OUTLOOK®; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. OUTLOOK®; PUMP, INFUSION Back to Search Results
Model Number 621-400ES
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device involved has been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As per the user facility: the pump had a burning smell with overheating.The top case was burned.No patient injury/involvement.The pump is operational but not repaired.There was no patient injury/involvement with this event.
 
Event or Problem Description
As per the user facility: the pump had a burning smell with overheating.The top case was burned.No patient injury/involvement.The pump is operational but not repaired.There was no patient injury/involvement with this event.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device involved has been received for evaluation.The reported complaint of thermal damage was confirmed.The power supply had evidence of resistive current path between 16v and the ground connections that increased in temperature until the immediate area of the power supply board/connectors and case top were damaged.The power supply and the case top were replaced and preventive maintenance service was performed.If additional pertinent information becomes available a follow-up report will be filed.If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OUTLOOK®
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
1601 wallace drive, suite 150
carrollton TX 75006 6690
MDR Report Key7613028
Report Number1641965-2018-00008
Device Sequence Number1946537
Product Code FRN
UDI-Device Identifier04046964186547
UDI-Public(01)04046964186547
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K011975
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number621-400ES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/05/2018
Supplement Date Received by Manufacturer06/05/2018
Initial Report FDA Received Date06/19/2018
Supplement Report FDA Received Date07/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-