• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-RAD MEDICAL DIAGNOSTICS GMBH REAGENT RED BLOOD CELLS BIOTESTCELL-I11

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-RAD MEDICAL DIAGNOSTICS GMBH REAGENT RED BLOOD CELLS BIOTESTCELL-I11 Back to Search Results
Catalog Number 816021100
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This is our combined initial and report on this incident.
 
Event or Problem Description
The customer reported discrepant results between immucor echo and tango infinity.Customer ran a patient sample that was 4+ pos for anti-jka on two different immucor analyzers and yielded a positive result.Testing the same tube was run on tango infinity with biotestcell 3, # 8815021-00 and biotestcell-11, #8817011-00 and the results were negative.The same sample was run on a second tango infinity and the result were also negative.The customer did neither provide the patient samples had caused discrepant results for investigational testing nor the allegedly defective lot of biotestcell-i11.Therefore our quality control laboratory tested their retained sample of biotestcell-i11 with different samples and controls (e.G.: anti-d, anti-jka) on tango infinity.The reagents that were used for testing on customers site were also used for qc testing.All test results were as expected, all positive reactions were clearly positive, no irregularities were found.Testing by our quality control laboratory confirmed that the allegedly defective lot of biotestcell-i11 functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.The two involved tango infinity's were inspected by our field service engineers.Both tangos were newly installed at the customer's site.The log files did not show any issue relevant abnormalities.The positive and negative solidscreen ii quality control resulted as expected.Other samples with positive antibody screen and panel results from the lab journal meet the criteria for the positive result evaluation.The instrument is working within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAGENT RED BLOOD CELLS BIOTESTCELL-I11
Common Device Name
BIOTESTCELL-I11
Manufacturer (Section D)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM  63303
MDR Report Key7613032
Report Number9610824-2018-00031
Device Sequence Number2221793
Product Code KSZ
UDI-Device Identifier07611969952434
UDI-Public(01)07611969952434(17)180618(10)8817011-00
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2018
Device Catalogue Number816021100
Device Lot Number8817011-00
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2018
Initial Report FDA Received Date06/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-