| Model Number |
10052 |
| Medical Device Problem Code |
Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: visual analysis of the returned cannula shows that the tip has separated from the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use while doing the routine count check, the scrub nurse discovered the soft tip of the cannula had dropped off.It was found within the patient's chest wall.The product was replaced.There was no adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: the complaint was confirmed for the tip coming off.It is unknown what may have caused this occurrence.The device history record was reviewed, the review found four individuals in training on the lot, with two performing the tip bonding process.The product passed all required inspection checks, with no non-conformances noted.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for similar future events.If information is provided in the future, a supplemental report will be issued.
|
| |
|