| Model Number |
18875 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device cuff started to leak after few hours.It was reported that it was tested without incident.It was reported that re-cannulation of a replacement tube was unknown.The customer reported that patient outcome was asked but unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample of endotracheal tube was received for evaluation.A visual inspection was performed and it was observed all the components are assembled properly at the tube.An inflation/deflation test was performed using a syringe, 25cc of air was applied to the cuff and it was noticed the cuff held the air, the cuff was left inflated 2 hours according to process instruction.After this time no deflation was observed, additionally the sample was submerged into a container filled with water and no leaks were observed.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|