| Model Number |
BIO-COMP SWVLK SP, 4.75X24.5MM |
| Health Effect - Clinical Codes |
Failure of Implant (1924); Rupture (2208)
|
| Date of Event |
05/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a rupture of the supraspinatus tendon the swivelock broke off when turning in.***update smer 12 june 2018: the surgery was completed successfully with a new device of the same part number.The new device was placed on a different spot - it is not covering the broken anchor.The anchor broke and 2/3 of the anchor will remain inside the bone well fixed.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation of the returned ar-2324bcm revealed that the swivelock screw had fragmented at the sixth distal thread.The most likely cause(s) of this type of event include advancing the implant before the driver tip has been driven to the proper depth or prying and/or leveraging the driver.
|
| |
|