Catalog Number 2C6671
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the filter of an interlink solution administration set filled with air and leaked total parenteral nutrition (tpn) from the end of the tubing.This event occurred while priming the device.It was further reported that once the device leaked it was then reprimed and the filter was adjusted.The device leaked tpn solution again.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A priming test was performed and the device passed with no issues noted.Pressure testing and clear passage testing under water were performed, and the device did not leak.Functional testing was performed on the filter, and the filter passed the testing with no issues noted.The reported condition was not verified.The device was found to operate per specification.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK SOLUTION ADMINISTRATION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
MDR Report Key7614277
Report Number1416980-2018-03711
Device Sequence Number19406207
Product Code FPA
UDI-Device Identifier00085412003993
UDI-Public(01)00085412003993
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K153158
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2023
Device Catalogue Number2C6671
Device Lot NumberR18B18032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2018
Initial Date Received by Manufacturer 05/23/2018
Supplement Date Received by Manufacturer07/09/2018
Initial Report FDA Received Date06/19/2018
Supplement Report FDA Received Date07/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-