MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0450
Device Problem Material Distortion (2977)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the port connector on recorder was very loose and the recorder was not connecting to the computer.There was patient information on the recorder that needs to be retrieved.The recorder has been returned for data extraction, but it failed as recorder could not be detected.The recorder did show a solid green led indicating that there was a study on the recorder.There was no patient and user harm, but a repeat procedure will need to be performed.
 
Manufacturer Narrative
Additional information: (evaluation summary): the bravo recorder was received for evaluation.The visual inspection found the usb connector was damaged.The investigation identified the root cause of the reported event to be associated to mishandling.This was determined to be a user error.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Corrected information: if information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
5920 longbow drive
boulder, CO 80301
7632104064
MDR Report Key7614353
MDR Text Key111840946
Report Number9710107-2018-00775
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101363064
UDI-Public7290101363064
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-0450
Device Catalogue NumberFGS-0450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-