| Model Number |
44420 |
| Medical Device Problem Codes |
Device Operates Differently Than Expected (2913); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the reservoir did not function well.The device was never implanted.The patient's status at the time of the report was alive-no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent and met the requirements for reflux, pressure-flow, pre-implantation, and leak testing.All valves are 100% tested at the time of manufacture.Due to imdrf harmonization, the device, method, result, and conclusion codes have been updated.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the reservoir rebounded slowly during pre-operative testing.The cause was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|